It is possible to make an accurate frequency measurement with a Frequency Counter of spectrum analyzer option.
Application
For example, as a test of wireless equipment of technical standards conformity, there is a "Frequency deviation ".
The meter to be used for the test is required 10 times or more accuracy than the allowable deviation of wireless equipment rule.
Solution
A Frequency Counter of spectrum analyzer MSA300 or MSA400 series option is used to make an accurate frequency measurement of technical standards conformity test. When using the Frequency Counter, the measurement accuracy is +/- 2ppm ( 23°C ).
Frequency deviation " technical standards conformity test of wireless equipment is executed with non-modulated continuous signal.
A Signal Generator non-modulated continuous signal is measured with spectrum analyzer MSA438, and the measurement results are compared between "with Frequency Counter" and "without Frequency Counter".
(Measuring the 2400MHz signal from Signal Generator)
System configuration
Spectrum Analyzer [MSA300 series, MSA400 series] | x 1 |
Frequency Counter(Factory option) | x 1 |
Products introduction
Handheld Spectrum Analyzer MSA400 Series
MSA300 series was extremely upgraded to MSA400 series.This product has large color display.
Handheld Spectrum Analyzer MSA300 Series
The best-seller model. The smallest size and the lowest price in the world.
Please feel free to contact us.
If you want to verify 5G, customize a radio wave shield box, or need product repair, please do not hesitate to contact us about any small matter.