Handheld spectrum analyzer enables easy C/N ratio measurement.
Judgement of stable receiving of such as digital broadcasting is made easy.

Application

In communication by digital modulation such as digital broadcasting, the carrier power vs noise ratio (C / N ratio) is important for stable receiving. Because deterioration of C / N ratio causes a rapid deterioration of the bit error rate (BER). As a result, it becomes impossible to reproduce the transmitted data correctly.

In our handheld spectrum analyzer MSA438, C / N ratio can be measured using “Channel power measurement” function.

  • Precise measurement and evaluation by a full-fledged spectrum analyzer.
  • Writing report becomes easy by saving measurement result (screen, data) into USB memory.
  • It is easy to use in the field because of small & lightweight equipment and 4 hours battery operation.

Solution

  • C / N ratio is expressed by the following equation on condition of carrier power Pc and noise power Pn.
  • So, both power values of a carrier in the specified band and a noise out of band are needed.
  • On the other hand, the handheld spectrum analyzer MSA438 has a channel power measurement function. It is possible to determine the C / N ratio from the following calculation measuring each of the channel power (carrier) and noise power.

A.Channel power(mW) = 10{Pc(dBm)/10}
B.Noise power(mW) = 10{Pt(dBm)/10} - Channel power(mW)

Pc: channel power in specified band / Pt: total power in screen
C/N ratio(dB) = 10*Log(channel power/noise power)

Handheld Spectrum Analyzer MSA438

The most popular model in MSA 400 series.
Measurement frequency : 50kHz to 3.3GHz

How to measure C/N ratio

  1. Set Band measurement mode of channel power measurement.
    [MEAS] key => select “BAND” with F1[MODE]
  2. Set the frequency and band of carrier.
    Adjust Band part to band of carrier by using encoder.
    Start measurement.
  3. Set Total measurement mode of channel power measurement.
    Select “Total” by F1[MODE].
    Start measurement.
    The value at lower right of screen is total power in screen Pt.
  4. C/N ratio is calculated based on the equation shown above.
    Pc=3.64dBm、Pt=3.72dBm
    1. Channel power= 2.312mW、Noise power=2.355mW – 2.312mW = 0.043mW
    2. C/N ratio= 10*Log(2.312/0.043) = 17.3dB
  5. ∴ C/N ratio
  6. 17.3dB

System configuration

3.3GHz Spectrum analyzer [MSA438]
Dedicated battery [MB400]

Products introduction

Handheld Spectrum Analyzer MSA438

The most popular model in MSA 400 series.
Measurement frequency : 50kHz to 3.3GHz

Lithium-ion Battery MB400

Product Description 7.4V/5000mAh7.4V/5000mAh Applied model MSA400/500 Series Related postsFrequency measurement with a Frequency Counter of spectrum analyzer option 01/08/2015 Field Intensity Measurement on the Road by Mobile Signal Analyzer 01/05/2014 Measurement of […]

Please feel free to contact us.

If you want to verify 5G, customize a radio wave shield box, or need product repair, please do not hesitate to contact us about any small matter.