The noise radiated from CNE (comparison noise emitted) widely used for the calibration such as anechoic chamber was measured with both of MR2300 and anechoic chamber, and the measurement results was compared.
- The right photo shows the state in which CNE is set up in MR2300 (Shield box:MY5310). The frequency bandwidth of the radiation noise was from 30MHz to 1GHz and the measurement was done at the vertical mode.
Measurement result in MR2300
Screen of PC software MSA230 for EMI test
Measurement result in anechoic chamber
-
[Result]
The noise radiated from CNE was measured in both of MR2300 and anechoic chamber, and almost the same data were obtained . Therefore, it was confirmed that it was very effective to use MR2300 as a debugging tool for the EMI problem
Products introduction
Please feel free to contact us.
If you want to verify 5G, customize a radio wave shield box, or need product repair, please do not hesitate to contact us about any small matter.